Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa 7542: (1 year) CAL VER /5500 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 40 NUMBER OF LINES: 134 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- R P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 30 minutes 1.005 DISP Temperature: -20 - 50degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {DC VOLTAGE} 1.008 DISP Set the Function to VDC. 1.009 DISP Connect the 5500A to the UUT as follows: 1.009 DISP [32] 5500A NORMAL HI to UUT V-OHM 1.009 DISP [32] 5500A NORMAL LO to UUT COM 1.010 HEAD DC VOLTAGE: {250mV Range} 1.011 5500 250 250.00mV 0.05% 0.02U 2W 2.001 HEAD DC VOLTAGE: {2500mV Range} 2.002 5500 2500 2500.0mV 0.07% 0.2U 2W 3.001 HEAD DC VOLTAGE: {25V Range} 3.002 5500 25 25.000V 0.07% 0.002U 2W 4.001 HEAD DC VOLTAGE: {250V Range} 4.002 5500 250 250.00V 0.07% 0.02U 2W 5.001 HEAD DC VOLTAGE: {1000V Range} 5.002 5500 1000 1000.0V 0.07% 0.2U 2W 6.001 HEAD {AC VOLTAGE} 6.002 DISP Press the AC/DC button to select AC. 6.003 HEAD AC VOLTAGE: {250mV Range} 6.004 5500 250 250.00mV 0.5% 0.1U 1kH SI 2W 7.001 5500 250 250.00mV 1% 0.4U 20kH SI 2W 8.001 5500 250 250.00mV 3% 2U 50kH SI 2W 9.001 HEAD AC VOLTAGE: {2500mV Range} 9.002 5500 2500 2500.0mV 0.5% 1U 1kH SI 2W 10.001 5500 2500 2500.0mV 1% 4U 20kH SI 2W 11.001 5500 2500 2500.0mV 3% 20U 50kH SI 2W 12.001 HEAD AC VOLTAGE: {25V Range} 12.002 5500 25 25.000V 0.5% 0.01U 1kH SI 2W 13.001 5500 25 25.000V 1% 0.04U 20kH SI 2W 14.001 5500 25 25.000V 3% 0.2U 50kH SI 2W 15.001 HEAD AC VOLTAGE: {250V Range} 15.002 5500 250 250.00V 0.5% 0.1U 1kH SI 2W 16.001 5500 250 250.00V 1% 0.4U 20kH SI 2W 17.001 HEAD AC VOLTAGE: {750V Range} 17.002 5500 750 700.0V 1% 1U 400H SI 2W 18.001 HEAD {dBm} 18.002 DISP Set the Function to dBm. 18.002 DISP Set the Range to AUTO. 18.003 5500 -38.00D 0.5U 1kH SI 2W 19.001 5500 -38.00D 1U 10kH SI 2W 20.001 5500 -30.00D 0.15U 10kH SI 2W 21.001 5500 -30.00D 0.15U 30kH SI 2W 22.001 5500 20.00D 0.15U 5kH SI 2W 23.001 5500 20.00D 0.5U 20kH SI 2W 24.001 5500 20.00D 1.2U 50kH SI 2W 25.001 5500 50.00D 0.15U 500H SI 2W 26.001 HEAD {FREQUENCY} 26.002 DISP Set the Function to FREQ. 26.003 5500 90.000H 0.05% 1V SI 2W 27.001 5500 900.00H 0.05% 1V SI 2W 28.001 5500 9.0000kH 0.05% 1V SI 2W 29.001 5500 90.000kH 0.05% 1V SI 2W 30.001 HEAD {RESISTANCE} 30.002 HEAD RESISTANCE: {25 MOhm Range} 30.003 5500 25 25.000MZ 1% 0.003U 2W 31.001 HEAD RESISTANCE: {2500 kOhm Range} 31.002 5500 2500 2500.0kZ 0.5% 0.2U 2W 32.001 HEAD RESISTANCE: {250 kOhm Range} 32.002 5500 250 250.00kZ 0.06% 0.02U 2W 33.001 HEAD RESISTANCE: {25 kOhm Range} 33.002 DISP Connect the 5500A to the UUT as follows: 33.002 DISP [32] 5500A AUX HI to UUT V/OHM 33.002 DISP [32] 5500A AUX LO to UUT COM 33.002 DISP [32] 5500A NORMAL HI to UUT V/OHM 33.002 DISP [32] 5500A NORMAL LO to UUT COM 33.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 33.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 33.003 5500 25 25.000kZ 0.06% 0.002U CW 34.001 HEAD RESISTANCE: {2500 Ohm Range} 34.002 5500 2500 2500.0Z 0.06% 0.2U CW 35.001 HEAD RESISTANCE: {250 Ohm Range} 35.002 5500 250 250.00Z 0.06% 0.12U CW 36.001 HEAD {DC LOW CURRENT} 36.002 DISP Set the UUT to A(DC). 36.003 DISP Connect the 5500A to the UUT as follows: 36.003 DISP [32] 5500A AUX HI to UUT 250mA 36.003 DISP [32] 5500A AUX LO to UUT COM 36.004 HEAD DC LOW CURRENT: {250mA Range} 36.005 5500 250 250.00mA 0.25% 0.02U 2W 37.001 HEAD {AC LOW CURRENT} 37.002 DISP Press the AC/DC button to select AC. 37.003 HEAD AC LOW CURRENT: {250mA Range} 37.004 5500 250 250.00mA 0.75% 0.1U 1kH SI 2W 38.001 HEAD {AC HIGH CURRENT} 38.002 DISP Connect the 5500A to the UUT as follows: 38.002 DISP [32] 5500A AUX HI to UUT 10A 38.002 DISP [32] 5500A AUX LO to UUT COM 38.003 HEAD AC HIGH CURRENT: {10A Range} 38.004 5500 10 10.000A 2% 0.01U 1kH SI 2W 39.001 HEAD {DC HIGH CURRENT} 39.002 DISP Press the AC/DC button to select DC. 39.003 HEAD DC HIGH CURRENT: {10A Range} 39.004 5500 10 10.000A 1% 0.002U 2W 40.001 END